实验室纯水机不出水的原因有哪些


实验室纯水机

随着生产科技的发展,实验室纯水机的应用越来越广泛,但是大家在使用超纯水机的过程中常常会出现机器不出水的故障,那么,引起实验室超纯水机不出水的原因有哪些呢?

With the development of production technology, laboratory pure water machine is used more and more widely, but everyone in the use of ultra pure water machine in the process often fails, no water machine that causes laboratory ultra pure water machine water is not what?


系统不制水
System does not make water
1、可能是自来水没开或水压不够,系统会发出报警声,打开自来水或短接低压保护开关。短接可能会造成增压泵空转而损坏泵。
1, may be the tap water is not open or water pressure is not enough, the system will issue a warning sound, open the tap water or short connect low voltage protection switch. Short circuit may cause damage to the pump and booster pump idling.
2、可能是预处理滤芯堵塞,水通量减小或没水通过,更换预处理滤芯。
2, may be a pre filter filter clogging, water flux decreases or no water through, replace the pretreatment filter.
3、低压保护开关损坏,不能正常闭合,更换低压保护开关。
3, low voltage protection switch is damaged, can not be normally closed, the replacement of low voltage protection switch.
4、高压保护开关损坏,不能正常闭合,此时满水指示灯会亮,更换高压保护开关。
4, high voltage protection switch is damaged, not close properly, then full of water LED lights to replace the high voltage protection switch.
系统一直造水:高压保护开关损坏,不能正常断开,制水指示灯一直亮,更换高压保护开关。
System has been made water: high voltage protection switch is damaged, can not normally open, water indicator light has been lit, the replacement of high voltage protection switch.
系统制水、满水转换太频繁:可能是纯水桶球阀未打开,不能储水,打开纯水桶球阀。
System to make water, full water conversion is too frequent: may be pure water bucket ball valve is not open, can not store water, open a pure bucket ball valve.


系统造水量偏小

System water consumption is small
1、可能是反渗透膜堵塞,更换反渗透膜。
1, may be the reverse osmosis membrane plug, replace the reverse osmosis membrane.
2、可能是供水不足,改善供水或更换预处理滤芯。
2, may be inadequate water supply, improve water supply or replace the pretreatment filter.
3、可能是增压泵损坏,更换增压泵。
3, may be damaged to the booster pump, replace the booster pump.


三级产水水质差:可能反渗透膜被损坏,更换反渗透膜。

Three water quality is poor: may reverse osmosis membrane is damaged, replace the reverse osmosis membrane.
一(三)级取水量小: 实验室超纯水机,超纯水机
One (three) level of water quantity is small: the laboratory ultra pure water machine, ultra pure water machine
可能是(超)纯化柱失效,更换(超)纯化柱。
May be (ultra) purification column failure, replacement (super) purification column.
可能循环管路上的单向阀损坏,造成水没有通过(超)纯化柱直接通过循环管路反向出水,更换循环管路上的单向阀。
May be caused by a one-way valve on the pipeline valve damage, resulting in water is not through the (super) purification column directly through the circulating pipeline reverse water, replace the one-way valve on the loop line.
可能纯水箱内水太少,等系统满水后再取水。
May be too little water and pure water tank, filled with water after the water intake system.
可能是纯水箱气囊内压力过低或无压力,重新给纯水箱气囊充气。
May be pure water tank air bag pressure is too low or no pressure, to pure water tank air bag inflation.
一级出水水质差
First grade effluent water quality difference
可能沉积水太久,将其排掉。
May deposit the water too long, it will be discharged.
可能循环泵损坏,不能进行循环处理,更换循环泵。
May be damaged by the pump, can not be carried out recycling, replace the circulating pump.
系统功能不正常:可能是系统电路损坏,更换电路板。 实验室超纯水机,超纯水机
System function is not normal: the system may be damaged, replace the circuit board. Laboratory ultra pure water machine, ultra pure water machine
一(三)级不出水:可能是一(三)级取水电磁损坏,更换一(三)级取水电磁阀。
One (three) level does not water: may be a (three) level of water damage, the replacement of a (three) level of water intake solenoid valve.
系统排放不出水:可能是系统排放电磁损坏,更换系统排放电磁阀。
System discharge is not water: may be the system discharge electromagnetic damage, the replacement of the system discharge solenoid valve.
系统冲洗不出水:可能是系统冲洗电磁损坏,更换系统冲洗电磁阀。
The system does not wash water: the system may be damaged, the replacement of the electromagnetic valve replacement system.
以上就是有关实验室纯水机不出水的故障解决办法,如果还不能解决您的问题,可以咨询产品的售后工程师,帮您尽快解决您的疑问。
The above is about the laboratory pure water machine does not water the fault solution, if you can not solve your problem, you can consult the product engineer, to help you solve your question as soon as possible.